Section 1 Introduction................................................................................................................1
Intended Use.............................................................................................................. 1
Section 2 Analyzer Components...............................................................................................2
Analyzer Component Illustrations............................................................................... 2
Section 3 Initial Startup..............................................................................................................3
Section 4 Analyzer Setup...........................................................................................................5
Test Settings ..........................................................................................................5
Type of Strip............................................................................................................... 5
Mode.......................................................................................................................... 6
Test Number............................................................................................................... 6
Continuous Test ......................................................................................................... 7
Operator ID ................................................................................................................ 7
Units...........................................................................................................................7
Analyzer Configuration.........................................................................................7
Printer Setup.............................................................................................................. 7
Sound......................................................................................................................... 8
Barcode reader .......................................................................................................... 9
Language................................................................................................................... 9
Date/Time................................................................................................................... 9
Auto Number Reset.................................................................................................. 10
Database...............................................................................................................10
Clear all Records.......................................................................................................11
Locate Record...........................................................................................................11
View Current Record.................................................................................................11
Self Test................................................................................................................12
QC Test.................................................................................................................12
QC Setup ................................................................................................................. 12
Run QC Test............................................................................................................. 13
Setting the QC Values.............................................................................................. 14
User Login............................................................................................................14
ID Admin .................................................................................................................. 14
Section 5 Analyzer Operation..................................................................................................17
Entering Canister Code ......................................................................................19
Normal Operation, no Barcode reader ..............................................................20
Sample/Strip Preparation......................................................................................... 20
Strip Test – Single Test Mode................................................................................... 20
Strip Test – Continuous Test Mode........................................................................... 22
Normal Operation, Barcode reader installed....................................................23
Strip Test – Single Test Mode with Barcode reader.................................................. 23
Strip Test – Continuous Test Mode with Barcode reader.......................................... 24
Urine Controls QC Testing .................................................................................26
Strip Preparation ...................................................................................................... 26
Urine Control Test Procedures................................................................................. 27
QC lockout ...........................................................................................................27
Section 6 Data/Communication...............................................................................................29
External Printer ........................................................................................................ 29
Barcode reader ........................................................................................................ 29
Section 7 Quality Control.........................................................................................................30
Section 8 Maintenance .............................................................................................................31
Loading Printer Paper.........................................................................................31
General Cleaning.................................................................................................32
Daily Cleaning......................................................................................................32
Remove the Strip Holder.......................................................................................... 32
Clean the Strip Holder.............................................................................................. 32
Load the Strip Holder ............................................................................................... 32
Return the Strip Holder/Mount to its Internal Position............................................... 33
Sample Deposit Cleaning ...................................................................................33
Strip Holder Sterilization.....................................................................................33
Section 9 Precautions ..............................................................................................................34
Section 10 Troubleshooting.....................................................................................................35
Appendix 1 Urine Analyzer Specifications.............................................................................36
Appendix 2 Compatible Urinalysis Reagent Strips................................................................37
Appendix 3 Urinalysis Reagent Strips Parameter Table .......................................................43
Appendix 4 Result Print-Out....................................................................................................44
Appendix 5 Barcode reader .....................................................................................................45
Appendix 6 Catalog ..................................................................................................................46
Appendix 7 Index of Symbols..................................................................................................47
Appendix 8 Warranty................................................................................................................48

Table of Contents

Questions and Answers

  • R
    Robert RichardsonAug 18, 2025
    What to do if the printer does not work on my Fisherbrand Measuring Instruments?
    • D
      Darrell TaylorAug 18, 2025
      If the printer on your Fisherbrand Measuring Instruments isn't working, first ensure that Auto-Print is enabled. Check if paper needs to be loaded and that the Printer Setup is set to Internal when an external printer is not in use. Make sure the paper roll is correctly placed, as incorrect placement will prevent printing. Clean the printer roller with ethanol and verify that the thermal paper is inserted with the proper side facing inward. Finally, check for and remove any paper jams by lifting the green paper release lever.
  • B
    Brian PerezAug 20, 2025
    Why is there no display on the screen of my Fisherbrand Measuring Instruments?
    • J
      Justin BurnsAug 20, 2025
      If your Fisherbrand Measuring Instruments shows no display on the screen, begin by turning the power switch on. Ensure that the analyzer is receiving power. Inspect the fuse located at the back of the analyzer for any signs of damage; replace it if necessary.
  • L
    Lauren WilliamsAug 23, 2025
    What to do if the optical sensor failed on my Fisherbrand Measuring Instruments?
    • E
      Eric MurphyAug 23, 2025
      If the Optical Sensor on your Fisherbrand Measuring Instruments has failed, make sure the Strip Holder is correctly locked onto the Strip Holder Mount, ensuring it snaps into place. Check that the white Calibration Circle is clean. Clean the Strip Holder. If problems persist, turn the power switch off and then on, and perform a Self-Test.
  • L
    lhayesAug 26, 2025
    What to do if I get a strip error on my Fisherbrand Measuring Instruments?
    • T
      Teresa MorenoAug 26, 2025
      If you encounter a 'Strip Error' on your Fisherbrand Measuring Instruments, verify that the type of strip used matches the selected strip setting. Also, confirm that the strip brand is compatible with the analyzer. Ensure that all test pads on the strip have been fully immersed in the specimen.
  • J
    Jared ManningAug 28, 2025
    What does excess light failed mean on Fisherbrand Measuring Instruments?
    • T
      tammyscottAug 28, 2025
      If your Fisherbrand Measuring Instruments indicates 'Excess Light Failed', make sure the analyzer is not exposed to direct light. Ensure that the white calibration circle is clean. Check that the Strip Holder is correctly positioned on the Strip Holder Mount. If the issue persists, try turning the power off and then on, and perform a Self-Test.
  • P
    pruittchristopherAug 30, 2025
    How to fix main control unit failed on Fisherbrand Measuring Instruments?
    • T
      taylorjenniferAug 30, 2025
      If the Main Control Unit of your Fisherbrand Measuring Instruments has failed, turn the power switch off, and then on again. Then, perform a Self-Test.
  • S
    stephanie07Sep 1, 2025
    How to replace a damaged fuse in Fisherbrand Measuring Instruments?
    • H
      Heather CookSep 2, 2025
      If the fuse in your Fisherbrand Measuring Instruments is damaged, turn the power switch off, disconnect the plug, and replace the damaged fuse with a new 2.0A fuse. A spare fuse is located in the fuse compartment.
  • K
    Kimberly PhillipsSep 4, 2025
    What causes test accuracy to fail on Fisherbrand Measuring Instruments?
    • N
      Nicholas FowlerSep 4, 2025
      If your Fisherbrand Measuring Instruments reports 'Test Accuracy Failed', turn the power switch off and then on again. Then, perform a Self-Test.
  • S
    Steven StewartSep 7, 2025
    What to do if the mechanism failed on my Fisherbrand Measuring Instruments?
    • D
      David WoodsSep 7, 2025
      If the Mechanism of your Fisherbrand Measuring Instruments has failed, first remove any obstructions blocking the Strip Holder's path. Avoid touching the Strip Holder while it is moving. If the problem persists, turn the power switch off and then on, and perform a Self-Test.
  • E
    estradadanielSep 9, 2025
    How to resolve a canister code error on Fisherbrand Measuring Instruments?
    • R
      robinsonfrankSep 9, 2025
      If you are experiencing a 'Canister Code Error' on your Fisherbrand Measuring Instruments, ensure that the canister code you entered is correct. Also, verify that the type of strip being used matches the analyzer setting.

Summary

Section 3 Initial Startup

Section 4 Analyzer Setup

Test Settings

Configures parameters like strip type, mode, test number, operator ID, and units.

Analyzer Configuration

Configures analyzer settings like printer, sound, barcode reader, language, and date/time.

Printer Settings Configuration

Configures printer parameters including darkness, mode, auto-print, and copies.

QC Test and Setup

Accesses Quality Control setup and running procedures.

Setting the QC Values

Enters target values for QC tests via a passcode protected screen.

User Login and ID Administration

Manages user access, operator ID, and administrator functions.

Section 5 Analyzer Operation

User Login and Initial Operation

Describes the login process and initial screen display based on user login status.

Normal Operation without Barcode Reader

Procedures for running the analyzer without a barcode reader.

Strip Test - Single Test Mode

Guides through performing a single strip test without a barcode reader.

Strip Test - Continuous Test Mode

Guides through performing multiple strip tests consecutively.

Normal Operation with Barcode Reader

Procedures for running the analyzer with a barcode reader.

Strip Test - Single Test Mode with Barcode Reader

Describes single strip testing procedures when a barcode reader is used.

Strip Test - Continuous Test Mode with Barcode Reader

Explains batch testing using barcodes for sample identification.

Urine Controls QC Testing

Outlines the procedure for performing quality control tests using control samples.

QC Lockout Notification

Explains the notification process when QC tests are required due to lockout.

Section 7 Quality Control

Quality Control Procedures

Outlines when and how to perform quality control checks for accurate results.

Section 8 Maintenance

Section 9 Precautions

Section 10 Troubleshooting

Troubleshooting Common Problems

A table listing common problems and their corresponding solutions for the analyzer.

Fisherbrand AimStrip Specifications

General IconGeneral
BrandFisherbrand
ModelAimStrip
CategoryMeasuring Instruments
LanguageEnglish