SD700-EU-EN V2.2 10/12
6
Advanced Settings
The SET function is used to accomplish the following:
• Format the SD memory card
• Set the date and time
• Set the sampling time
• Set the beeper sound ON/OFF
• Set the SD card decimal/comma character
• Select the units of measure for temperature and barometric pressure
• Set the RS232 data output ON/OFF
Note: The datalogger function must be switched OFF before attempting to enter the Advanced
Settings mode.
Press and Hold the SET button for >2 seconds. The first function (Sd F) will appear in the display.
Press the SET button to step through all of the functions. Use the ▲ and ▼ buttons to adjust the
selected function. Use the “LOG” button to step through fields within a function. In the SET mode, if
no button is pressed within 5 seconds the logger will revert to the standard mode.
1. Sd F: Format the SD card. Press the ▲ button to select yES or no. For yES, press the Enter
button to format the card and erase all existing data.
2. dAtE: Set the date and time. Press the ▲ or ▼ buttons to adjust the selected field. Press the
Enter button to store the value and to step through the various fields.
3. SP-t: Set the sample rate. Press the ▲ button to select the desired sample rate and press
Enter to store the selection. The selections are: 5, 10, 30, 60, 120, 300, 600 seconds and
AUTO. In AUTO, the data will be stored every time there is a value change of >1% RH OR
>1°C
4. bEEP: Set the beeper ON or OFF. Press the ▲ button to select ON or OFF and press Enter to
store the selection.
5. dEC: Set the SD card decimal character. Press the ▲ button to select USA (decimal) or Euro
(comma) and press Enter to store the selection.
6. t-CF: Set the Temp. unit to °F or °C and press Enter to store the selection.
7. rS232: Set the RS232 data output ON/OFF. Press the ▲ button to select ON or OFF and
press Enter to store the selection.
8. Baro: Set the unit of measure for barometric pressure. Use the ▲ button to select inches of
mercury (InHg), millimeters of mercury (-Hg), or hectoPascal (hPa).
9. ESC: Exit the setting mode. Press the SET button to return to normal operation.
System RESET
If a condition appears where the CPU does not respond to keystrokes or the meter seems frozen,
press the RESET button on the side of the datalogger (use a paper clip or similar pointed object) to
return the meter to a working state.
RS232 Interface
A serial output is provided to link the meter to a PC’s serial port. Call Technical Assistance for more
information on this interface.