ProSim™ 8
Getting Started Manual
22
Leak Test
Target Pressure .................................................. 20 mmHg to 400 mmHg
Elapsed time ....................................................... 0:30 minutes:seconds to 5:00 minutes:seconds in 30 second steps
Range ................................................................. 0 mmHg/minute to 200 mmHg/minute
Internal Leak rate .................................................... <2 mmHg/min into 500 ml rigid volume
Pressure Relief Test Range ................................. 100 mmHg to 400 mmHg
Oximeter SpO2 Optical Emitter and Detector (optional)
% O
2
Range ................................................................. 30 % to 100 %
Resolution ........................................................... 1 %
Accuracy
With oximeter manufacturer’s R-curve
Saturation within UUT specific range .......... ±(1 count + specified accuracy of the UUT)
Saturation outside UUT specific range ....... monotonic with unspecified accuracy
With Fluke Biomedical R-curves
91 % to 100 % ............................................ ±(3 counts + specified accuracy of the UUT)
81 % to 90 % .............................................. ±(5 counts + specified accuracy of the UUT)
71 % to 80 % .............................................. ±(7 counts + specified accuracy of the UUT)
Below 71 % ................................................. monotonic with unspecified accuracy
Heart Rate
Range ................................................................. 30 BPM to 300 BPM in 1 BPM steps. Oximeter SpO2 Optical Emitter and Detector is
synchronized with ECG rate delayed by 150 ms.
Accuracy ............................................................. ±1 % of setting