5220A
Instruction Manual
2-10
7. Determine the active output terminals (front or rear) by connecting a short jumper
wire between the front-panel CURRENT OUTPUT terminals and then pressing the
OPR/STDBY switch. If the OVERCOMPLIANCE LED lights, the rear output
terminals are active. Otherwise, the OPR LED will light indicating that the front
CURRENT OUTPUT terminals are active.
8. Press the STDBY switch and then remove the short from the front output terminals.
9. Connect the intended load between the active output terminals.
10. Adjust the control voltage to the desired level.
11. Command the operate mode. If the OPR LED is lit, the predetermined current level is
flowing through the load.
12. Set the unit to standby before disconnecting the load or setting the POWER switch to
OFF.