The
instrument
parameters at
power-on are listed in
Table
2-21
. The remote clear
commands
can
be used to
reset all
parameters
except the last
memory location and the remote/ local state.
Table
2-19. Command-Parsing
Time
COMMAND COMPONENT
TIME
Header
Boolean
Numeric
Unsigned
Integer
Numeric
Floating-Point
Numeric
Trigger-String
Numeric
Suffix
Learn-Interf
ace
Suffix
2
ms
1 ms
2 ms
+
1 ms
per
character
2
ms
+
1 ms per
character
10
ms
+
0.5 ms per
character
1
to 1 .5 ms
40 ^ms
Table
2-20.
Typical
Programming Time of the
Generator Functions
1
FUNCTION
TIME
(IN ms)
NOTES
Frequency
70
1, 2, 3,
4
Amplitude in
Volts 60 1,
2, 4,
5
Amplitude
in dBm 95
1, 2, 4,
5
AM Depth
50
2.
4
FM
Deviation
35
4
Modulation
Frequency 25
Enable/Disable
AM
55
2,
Enable/Di
sable FM
25
Recall 98
190
6,
5
RF
Output
On 50
2
RF
Output
Off 35
Notes
1.
May take up
to
5 ms
longer if
the
relative
mode is
enabled,
2. Can save
10 ms if
all
Level
correction
is
disabled
with
special
function
81
.
3. Add
20 ms
when
frequency
changes
from
greater
than or
equal
to
245
MHz
to less
than
245 MHz.
4. Edits
and
steps
may take up
to 5
ms
longer
than
the
programming
function
directly,
5.
Add 15 ms
when
the
Attenuator
settings
change,
6.
Recalls
vary
considerably
depending
on
the
stored
data.
Maximum
is
approximately
250 ms.