Table
2-21.
IEEE-488 Power-On State
INSTRUMENT PARAMETER
STATE
NOTES
Memory
location
parameters
same
as power
off
1
RF on/off
same
as power off
Last memory
Location
same
as power off 2
Remote/Local
state
local
IEEE
output buffer
cleared
IEEE input buffer
cleared
Valid mode
same
as power off
Record
terminator
same
as power
off
Unbuffered mode
same as power
off
Record mode
same
as power off
SRQ mask
same
as
power
off
Trigger
configuration
same
as power off
SRQ
interface
signal
may
be asserted 3
Notes
1. The
contents of memory
location 98 (Instrcmient Preset State) is
listed in
Table
2-3.
2.
The
last memory location
is used for sequence operations.
3,
If
the
SRQ
mask has the
power on, output valid, or ready 1>RQ enabled.
the SRQ interface signal is asserted
on
power-on.