Performance characteristics
Accuracy
Within ±15 mg/dL (0.83 mmol/L) at glucose concentration < 75 mg/dL
(4.2 mmol/L) and within ±20% at glucose concentration ≥ 75 mg/dL (4.2
mmol/L).
Precision
CVs (%) of intermediate precision and repeatability are less than 5%.
The product has been tested to meet the requirements of ISO 15197.