Forma Scientific, Inc.___________________________________________________Operation
4-4
h. Receptacle Reset Button (15 Amp)
The receptacle reset button (located directly above the Blower Motor/Lights Reset
Button) is an inline circuit breaker for the receptacles only. If an overload occurs the circuit
breaker will trip, and the button will protrude from the panel. Depressing the button resets the
circuit breaker.
1. Turn the power off.
2. Unplug the unit.
3. Turn the receptacle power off.
4. Press the receptacle reset button.
i. Duplex Receptacles
Duplex receptacles (115 volts, 15 amps) are located on the left and right side wall of the
work station. Power is controlled by the receptacle switch located on the control panel.
j. Drain Valve
The drain valve, located on the right front side of the cabinet, is provided for the safe
drainage of the drain pan. This valve must remain closed while work is being performed in the
cabinet and be used only in the event of a major spill.
CAUTION! If a spill occurs, immediately consult a biological safety officer or other
qualified individual for proper procedures. To contain a spill, connect a sealed hose from the
drain valve to a sealed container.
k. Service Valves
Two service valves are standard with each cabinet. These valves are located on the right
and left side of the work station and can be coded with the type of service that they supply.
Identification index buttons are supplied.
The cabinet supports six service valves. Additional valves may be purchased from
Forma Scientific Inc.
l. Exhaust Filter Guard
The exhaust filter guard, located on top of the exhaust filter, protects the exhaust air
flow and prevents the storing of objects on top of the housing.