Luftgehaltsprüfer 8 Liter
Air entrainment meter 8 liter
Seite
/
page
26
Readadjustment of the manometer’s scale!
Should the manometer pointer doesn’t show 5% after the calibration
process you have to readjust the manometers scale:
Shows the pointer e.g. 4%, remove the plastic plug on the back side of
the manometer (zero-point correction). Adjust the manometer’s pointer
with the enclosed square socket screw key, so that the manometer pointer
stand exactly on 5%. Put the plastic plug on again.
Repeat the calibration process!
Please clean and dry the equipment after every test!