Section 6 Parameter Settings
46 FT702LT Flat Front Wind Sensor Manual
6.4.13 ID: Set or Query the Listener and Talker Identifiers
Command
Parameter
ID
Command
Syntax
SET Sensor:
$<listenerID>,ID<RxID><TxID>*<checksum><cr><lf>
$aa,ID=cccc*hh<cr><lf>
QUERY Sensor:
$<listenerID>,ID?*<checksum><cr><lf>
$aa,ID?*hh<cr><lf>
FT702LT output:
$<talkerID>,ID=<RxID><TxID>*<checksum><cr><lf>
$aa,ID=cccc*hh<cr><lf>
Parameters
<RxID>
00 to ZZ
The FT702LT 2 digit listener address identifier
(Factory Default RxID = 01)
<TxID>
00 to ZZ
The FT702LT 2 digit talker address identifier
(Factory Default TxID = WI)
Examples
Example 1
Set the FT702LT listener address identifier to A1 and the talker address identifier to B1.
Verify that the command has been accepted.
Message
Comment
$01,IDA1B1*//<cr><lf>
Set address ID’s
$A1,ID?*//<cr><lf>
Query ID settings
$B1,ID=A1B1*6C<cr><lf>
FT702LT response
Note: the ID? command must use the new listener ID otherwise the command will not be
recognised.
Description
Use the ID command to set the listener and talker address identifiers. See Section 5.3.4 for
information on listener and talker address identifiers.