Section 6 Parameter Settings
53 FT702LT Flat Front Wind Sensor Manual
6.4.20 UC.2: Clear User Calibration Table Record
Command
Parameter
UC (Erase table)
Command
Syntax
SET Sensor:
$<listenerID>,UC<erase>*<checksum><cr><lf>
$aa,UCCLEAR*hh<cr><lf>
Parameters
<erase>
CLEAR
Erases both Flash and RAM copies of user calibration table
Examples
Example 1
Erase the user calibration tables and verify
Message
Comment
$01,UCCLEAR*62<cr><lf>
Erase calibration tables
$01,UC?*04<cr><lf>
Query user calibration table status
$WI,UC=00,D,0000,0000*71<cr><lf>
FT702LT response
Description
Use the UCCLEAR command to erase the RAM and saved FLASH copies of the user
calibration table. A UCCLEAR is performed before a new user calibration table is loaded
(see section 6.4.21).
The user calibration table label is also cleared to 32 ASCII spaces when the UCCLEAR
command is sent (see section 6.4.23).
The User Calibration Table is only available from build version 22 and above.