3.11
3.12
Is Ottoman reclining signal line and Ottoman reclining driving
signal line (That of AS-870 is short pin.) inserted surely to the
connector (CN6、CN5、CN7)?
No
Connect connector surely.
Service No.「26」
Is there any broken line in the Ottoman moving unit signal line
(4 center core) ?
(Confirm by exchanging tester or stretching sensor unit of
confirmin
items(non-defective
roducts)(AS-1000 onl
)
No
Exchange sensor unit for Ottoman moving
ni
Exchange control basement board
es
es
Service No.「27」 (AS-1000 only)
es
Is the junction connector for Ottoman driving connected
surely? (AS-1000 only)
No
Connect connector surely
es
Is Ottoman driving 6P code for connection of control unit
broken ? (That of AS-870 is short pin)
No
es
Exchange Ottoman driving 6 code.
Exchange short pin for AS-870
Is Ottoman or reclining signal line broken ?
After exchange alternate actuator, is there any other error
occurred ?
No
Exchange concerned actuator.
Junction connector of motor 2 pin code and stretching sensor
signal 6 pin code for Ottoman moving unit is surely connected ?
No
Connect surely junction connector
Is Connector of each code for Ottoman moving unit surely
connected 6 pin connector (CN7) for sensor signal of control
basement board, and Motor 2 pin connector (CN13) ?
No
Connect surely connector
Is there any break line of control basement side 6 pin code for
Ottoman moving sensor signal, and 2P code for stretch
Exchange control basement side
concerned code
“About DC24V” is come up on the connector (CN13) of control
basement board when you push button 「lengthen」、「shorten」
of remote controller ?
No
Exchange control basement board
Is there any breakdown of Ottoman moving unit side sensor
basement board or break line of 4 center core code for sensor
signal ?
Exchange sensor assy (with harness )
Exchange Ottoman Stretch Moving Assy
es
es
No
es
No
es
es