What to do if Future Diagnostics Blood Glucose Meter test does not start after applying sample?
- SshannonhallAug 3, 2025
If the test does not start after applying the sample, it could be due to a defective test strip. In this case, repeat the test with a new test strip. Another reason may be that the sample was applied after the automatic switch-off (3 minutes after the last user action). Repeat the test with a new test strip, and apply the sample only when the flashing symbol appears on the display.