E - Display the internal raw counts.
F1.1 Digital Averages
Number of A/D conversions which comprise the running average.
Choices: 1, 2, 4, 8, 16, 32
F1.2 Display Update Rate
Number of display updates per second.
Choices: 20, 10, 5, 2, 1, 0.50, 0.25
F1.3 Motion Limit in Digits
Select the width of the motion window by use of the TARE key. A selection of d will
disable the no-motion requirement for use by the zero, tare and print functions.
Choices: 0.5, 1, 2, 3, 4, 5, 6, d
F1.4 Auto zero tracking Enable
d - AZT disabled.
E - AZT enabled.
F1.5 Auto zero tracking width
Number of digits over which zero will track.
Choices: 0.5, 1, 2, 3
F1.6 Two Percent Zero Range Enable
d - Zero active at any weight.
E - Zero active to +/-2% of capacity about zero.
F1.7 Key Functions Enable
d - All right-hand keys disabled.
E - All key functions enabled.