11
Irregular Heartbeat Detector
This symbol - indicates that certain pulse irregularities were detected during the measurement.
In this case, the result may deviate from your normal basal blood pressure – repeat the measurement.
Information for the physician on frequent appearance of the Irregular Heartbeat Symbol.
This instrument is an oscillometric blood pressure monitor device that also analyzes pulse frequency
during measurement. The instrument is clinically tested.
If pulse irregularities occur during measurement, the irregular heartbeat symbol is displayed after the
measurement. If the symbol appears more frequently (e.g. several times per week on measurements
performed daily) or if it suddenly appears more often than usual, we recommend the patient to seek
medical advice. The instrument does not replace a cardiac examination, but serves to detect pulse
irregularities at an early stage.
5.4 Error Indicates
The following symbol will appear on the display when measuring abnormal
No display appears
Weak battery or improper
placement
Replace both batteries with new ones.
Check the battery installation for proper
placement of the battery polarities.
Er 1 Sensor abnormal
Check if the pump is working or not. If it is
working, then the problem is sensor
abnormal. Please send it to the local
Er 2
Monitor could not detect pulse
wave or cannot calculate the
blood pressure data
Check if the air releasing is too slow or
not. If it is too slow, please check if there
is any dust in the tube plug of the cuff
and the cuff port in the device. If yes,
please clean and start the measurement
again. If no, please send the device back
to the local distributor.
Er 3
Measurement result is abnormal
(SYS≤35mmHg, DIA≤23mmHg)
Occasionally-measure for one more time
/ Always - send it to local distributor
Er 4
Too loose cuff or air leakage
(Cannot inflate to 30mmHg within
15s)
Tie the cuff correctly and make sure the
air plug is properly inserted in the unit
Er 5 The air tube is crimped
Correct it and make the measurement
again
Er 6
The sensor is sensing great
fluctuation in the pressure
Please keep quiet and don’t move
Er 7
The pressure that the sensor
sensing is over the limit
Please send back to the local distributor
Er 8
The demarcation is incorrect or
the device has not been
Please send back to the local distributor