8 Certifications
Device standard:
This device is manufactured to meet the European blood pressure monitors: EN1060-3 / IEC 80601-2-30
/ ISO81060-1 / IEC60601-1-11 / IEC60601-1
Electromagnetic compatibility:
Device fulfills the stipulations of the International standard IEC60601-1-2
The declaration of conformity is available on:
http://DOC.hesdo.com/FB160-DOC.pdf
9 Technical Specifications
Model: C02 / FB160
Weight: 326.4g(Batteries are included)
Display: 62.5 * 63.5 mm LED Digital Display
Size: 123 (W) x 82 (L) x 25 (H) mm
Accessories: 1×Main Device, 1×Cuff, 1×Users manual, 1x case, 3x AA batteries
Operating Conditions: Temperature: 5°C to 40°C; Humidity: 15% to 93% RH;
Storage And Shipping Conditions: Temperature: -25°C to 70°C; Humidity:≤ 93% RH;
Atmospheric pressure range:70kPa~106kPa
Measuring method: Oscillometric
Pressure sensor: Resistive
Measuring range: DIA: 40 – 130 mmHg SYS 60 - 230
Pulse: 40 to 170 per minute
Cuff pressure display range: <300mmHg
Memory: Automatically stores the last 120 measurements for 2 users (total 240)
Measuring resolution: 1 mmHg
Accuracy: Pressure within ± 3 mmHg / pulse ± 5 % of the reading
Power source: a) 3*AA batteries, 1.5 V
b) AC adapter INPUT: 100-240VAC 50/60HZ OUTPUT: 5V DC 1A
Accessories: Wide range rigid cuff 22 - 40 cm
Automatically power off: 60 seconds
Users: Adult
10 EMC Declaration
1) *This product needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided, and this unit can be affected by portable and
mobile RF communications equipment.
2) * Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This
may result in incorrect operation of the unit.
3) * Caution: This unit has been thoroughly tested and inspected to assure proper performance and
operation!
4) * Caution: this machine should not be used adjacent to or stacked with other equipment and that if