Guidance and manufacture’s declaration electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer
or
the user of device should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact ±2
kV, ±4 kV, ±8
kV, ±15 kV air
±8 kV contact ±2 kV,
±4 kV, ±8
kV, ±15 kV air
Floors should be wood, concrete
or ceramic tile. If floor are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC
61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines IEC
61000-4-11
0 % UT; 0.5 cycle
at 0°,45°,90°,
135°, 180°, 225°,
270°, 315°
0 % UT ; 1 cycle
70 % UT; 25/30
cycle
0% UT; 250/300
cycle
Not applicable
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the device requires continued
operation during power mains
interruptions, it is recommended
that the device be powered from
an uninterruptible power supply
or a battery.
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
50/60Hz
30 A/m
50/60Hz
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.