34
Connect the temperature probe to the Comms port at the top of the instrument.
Locate a suitable gland or bung on the pipework and insert the temperature probe.
When the probe is as far into the gland as possible without opening the tap, open the gas tap.
Finally, push the probe into the gland far enough to ensure that the tip is in the centre of the gas pipe.
Now press the “Sample” key to begin measuring the temperature in the pipe.
When the reading has stabilised, press the “Sample” key to retain the reading. This will return the
highlight area around the temperature figure. The temperature probe can now be removed from the
pipe.
Warning
Warning
Take great care when inserting the temperature probe to avoid gas
leakage.
Take great care when removing the temperature probe. In particular,
avoid standing with your face in direct line with the gland because gas
can be expelled at great force.