ApexProandCARESCAPETelemetry
4.Wipeoffthecleaningsolutionswithaclean,lightlymoistenedlint-freecloth.
5.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Allowthebatterycompartmenttoairdrycompletelypriortoclosingthe
compartmentdoor.
Dryingtimesmayvarybasedontheenvironmentalconditions.
6.Installthebatteryandclosebatterydoor.
CARESCAPETelemetryserver:HardwareseriesMP100R
Completethefollowingproceduretocleantheexteriorsurfacesofthetelemetry
server.
1.Shutdownthetelemetryserver.Formoreinformation,seethedocumentation
accompanyingthetelemetryserver.
2.Turnoffpowertothedevice.
3.Disconnectthedevicefromthepowersupply.
4.Removeallcables.
5.Dampenaclean,softlint-freeclothwithoneofthepermitteddetergents.
6.Wringexcessliquidfromtheclothandwipetheexteriorsurface.Payspecial
attentiontohard-to-cleanareas,likegroovesandcrevices.
Anycontactofcleaningsolutionswithmetalpartsmaycausecorrosion.
Donotdamageorbendconnectorpinswhencleaningordrying.
Donotletuidpoolaroundconnectionpins.Ifthishappens,blotdrywith
acottonswaborsoftcloth.
7.Wipeoffthedetergentswithaclean,lightlymoistenedcloth.
8.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Dryingtimesmayvarybasedontheenvironmentalconditions.
9.Reconnectthedevicetothepowersupply.
10.Reconnectallcables.
11.Turnonthepowertothedevice.
Internalcomponents
1.Checkthatthetelemetryserverispoweredoffandunplugged.
2.Disconnectallcords/cablesfromthetelemetryserver.
3.Removetheaccesspanelsonthetelemetryserver.Usingacompressedair
dusterrecommendedforcomputersandelectronics,carefullyblowoutany
accumulateddust.
Fans
Removeallofthedustfromthefansandthefanintakes.Thefansandfanintakes
usuallyaccumulatethemostdust.Accumulateddustcanclogthefansandfan
intakesandpreventpropercooling.
26
Cleaninganddisinfecting
2110145-001

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CARESCAPE MP100R and is the answer not in the manual?

GE CARESCAPE MP100R Specifications

General IconGeneral
BrandGE
ModelCARESCAPE MP100R
CategoryMedical Equipment
LanguageEnglish

Related product manuals