EasyManua.ls Logo

GE CareScape V100 - Cleaning Non-Applied Parts, General Instructions; Other Devices Cleaning Instructions; Disinfection; Disinfection Points to Note

GE CareScape V100
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CARESCAPEB850Monitor
Cleaningnon-appliedparts,generalinstructions
Followtheseinstructionstocleanthemonitor,modules,EEGheadbox,andother
non-appliedpartsunlessthereareseparatepart-specicinstructions.Non-applied
partsrefertothosepartsofthesystemthatarenotindirectcontactwiththepatient.
Beespeciallycarefulwhencleaningthedisplayscreen.Theyaremoresensitiveto
roughcleaningmethodsthanforexamplethemonitorhousing.
1.Turnoffthepowertotheequipment.
2.Disconnecttheequipmentfromthepowersupply.
3.Removeallcables.
4.Dampenasoftlint-freeclothwithoneofthepermitteddetergents.
5.Wringexcessliquidfromtheclothandwipetheexteriorsurface.Payspecial
attentiontohardtocleanareas,likegroovesandcrevices.
Anycontactofcleaningordisinfectingsolutionswithmetalpartsmaycause
corrosion.
Donotdamageorbendconnectorpinswhencleaningordrying.
6.Removeanysoilassoonaspossibleafterusebywipingthedevice.
Donotletuidpoolaroundconnectionpins.Ifthishappens,blotdrywithacotton
swaborsoftcloth.
7.Wipeoffthecleaningsolutionswithaclean,lightlymoistenedcloth.
8.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Dryingtimesmayvarybasedontheenvironmentalconditions.
9.Visuallyensurethatthedeviceisclean.Ifvisiblesoilremains,repeatcleaning
procedureuntilthedeviceisthoroughlyclean.
10.Reconnecttheequipmenttothepowersupply.
11.Turnonthepowertotheequipment.
Otherdevices’cleaninginstructions
Fordetailsaboutcleaningtheaccessories,seetheinstructionsforuseinthe
accessorypackage.
Donotreusesingle-usedisposableaccessories.
Refertotheuserdocumentationprovidedwiththekeyboard,mouse,andbarcode
readerforinstructionsonhowtocleanthem.Alwaysconsideryourhospitalguidelines
aswell.
Disinfection
Disinfectionpointstonote
Alwayscleanbeforedisinfecting.
Alwaysdilutecleaninganddisinfectantagentsaccordingtotheirmanufacturer’s
instructions.Alwaysconsideryourhospitalguidelinesaswell.
Useonlythepermittedsubstances.
2110145-001
Cleaninganddisinfecting
57

Table of Contents

Other manuals for GE CareScape V100