Electrical safety tests
Ultrasound SystemCommon Service Information 4-41
Direction 5444964-100 English Rev. 5
Isolated Probe (Sink) Leakage-Isolation Test (continued)
Record all data and keep the record of the result with other hard
copies of maintenance data.
NOTE: Not all test procedures are applicable to all areas of the world.
Reversed Polarity testing content satisfies regions following IEC
62353:2007 and IEC 60601-1:2005.
NOTE: Probe names in italics font are given as examples only.
Table 4-18: Typical data format for recording Isolated Probe (sink) leakage
Unit under test___________________________ Date of test:___________
Test Conditions Probe as measured in saline bath
System Power Grounding /PE 4C i12L TS E8C
off closed
on closed
Table 4-19: Typical data format for recording Isolated Probe (sink) leakage
Unit under test____________________________________ Date of test:_____________
Test Conditions Probe as measured in saline bath
System
Power
Grounding/
PE
Limit
µA 4C i12L TS E8C
off closed 5000
on closed 5000
off closed
(reversed
polarity)
5000
on closed
(reversed
polarity)
5000

Table of Contents

Other manuals for GE Vivid i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid i and is the answer not in the manual?

GE Vivid i Specifications

General IconGeneral
TypePortable Ultrasound System
Transducer Ports2 active ports
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Imaging Modes2D, M-Mode
ConnectivityDICOM, Ethernet, USB
ApplicationsVascular, Abdominal
WeightApproximately 5.5 kg (12.1 lbs)
Battery LifeUp to 1 hour (depending on usage)
ProbesPhased array, Linear array, Convex array

Related product manuals