Site Preparations
3-10 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Electrical requirements for Console Ultrasound Systems (continued)
Table 3-6: Electrical Requirements - Voluson Consoles
Product Voltage Tolerance
Current or
Power
Consumption
Frequency
Voluson® 730
100 VAC
±10%
10.1 A
50, 60 Hz (±2Hz)
110 VAC 9.2 A
115 VAC 8.8 A
130 VAC 7.8 A
230 VAC 4.4 A
240 VAC 4.2 A
Voluson E6 /
Voluson E8
model-type <=BT13
100 - 130 VAC 10.3 - 7.7 A
220 - 240 VAC 4.5 - 4.2 A
Voluson E6 /
Voluson E8
model-type >BT13
Voluson E10,
>BT15
100-240 VAC 800 VA
Voluson P6 /
Voluson P8 100-240 VAC 900 VA
Voluson S6 /
Voluson S8
100-130 VAC
200-240 VAC 900 VA
Voluson S10 100-120 VAC
220-240 VAC 900 VA

Table of Contents

Other manuals for GE Vivid i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid i and is the answer not in the manual?

GE Vivid i Specifications

General IconGeneral
TypePortable Ultrasound System
Transducer Ports2 active ports
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Imaging Modes2D, M-Mode
ConnectivityDICOM, Ethernet, USB
ApplicationsVascular, Abdominal
WeightApproximately 5.5 kg (12.1 lbs)
Battery LifeUp to 1 hour (depending on usage)
ProbesPhased array, Linear array, Convex array