Site Preparations
3-14 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Electrical Requirements for Isolation Cart
Table 3-11: Isolation Cart for LOGIQBook XP Series/LOGIQ e/LOGIQ e Vet/LOGIQ i/
Vivid e
Input Output
100-120V~, 50/60Hz, 500VA 100-120V~, 50/60Hz, 250VA
220-240V~, 50/60Hz, 500VA 220--240V~, 50/60Hz, 250VA
Table 3-12: E-Isolation Cart for LOGIQ e
Input Output
100-120V~, 60Hz, 400VA 100-120V~, 60Hz, 350VA
220-240V~, 50Hz, 400VA 220--240V~, 50Hz, 350VA
Table 3-13: Advanced Isolation Cart
Input Output
100-120V~, 60Hz, 400VA 100-120V~, 60Hz, 350VA
220-240V~, 50Hz, 400VA 220--240V~, 50Hz, 350VA

Table of Contents

Other manuals for GE Vivid i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid i and is the answer not in the manual?

GE Vivid i Specifications

General IconGeneral
TypePortable Ultrasound System
Transducer Ports2 active ports
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Imaging Modes2D, M-Mode
ConnectivityDICOM, Ethernet, USB
ApplicationsVascular, Abdominal
WeightApproximately 5.5 kg (12.1 lbs)
Battery LifeUp to 1 hour (depending on usage)
ProbesPhased array, Linear array, Convex array