VIII. UNIT CLEANING
CAUTION: To reduce the risk of an electric shock, do not remove the rear
cover of the device. Refer all servicing to qualied personnel.
CAUTION: Disconnect the unit power cord from the power receptacle
prior to performing cleaning operations.
Periodic external cleaning of the TCU is required. Clean all external
surfaces with a mild soap and water solution using a moist, soft, lint-free
cloth. Dry all surfaces with a clean, lint-free cloth.
CAUTION:
Care must be exercised to ensure that the electrical connections and con-
trol knob do not come in contact with water.
IX. STORAGE
Store the TCU in a cool, dry place.
X. STERILIZATION OF ACCESSORIES
The Cautery Handpiece and Reusable Cautery Electrodes may be auto-
claved according to the instructions below.
Separate the electrode from the Handpiece prior to autoclaving and use
aseptic procedures when handling and sterilizing both items.
1. Wipe and remove all debris present on the Handpiece/Electrode.
2. Place the Handpiece/Electrode in an autoclavable pouch or wrapped
cloth.
3. Steam sterilize for a period of 20 minutes at 250°F (121°C) or consult
an appropriate nationally recognized standard for sterilizing medical in-
strumentation. Do not ash autoclave or exceed 250°F.
4. Allow the Handpiece/Electrode to cool for 30 minutes prior to use.
5. Inspect the Handpiece/Electrode after sterilization for damage. If the
Handpiece/Electrode has deteriorated or cracked, or severe discoloration
or damage has occurred, discard and replace.
CAUTION: Do not use liquid sterilizing solutions with TCU or accessories,
as the potential for material damage exists.
5