22
A
measuring
range:
body mode:
object mode:
34.0 - 42.2 °C (93.2 - 108.0 °F)
0 - 100 °C (32 - 212 °F)
Resolution: 0.1 °C oder 0.1 °F
Accuracy
(laboratory):
36.0 °C - 39.0 °C: +/- 0.2 °C
(96.8 °F - 102.2 °F: +/- 0.4 °F);
other ranges +/- 2 %
Measuring
time:
approx. 1 second
Display: Liquid crystal display, 4 digits
Battery: 2 x 1.5 V, size AAA
Lifetime of
battery:
approx. 1000 measurements
Fever alarm: 10 short beeps and red LCD
backlight if the reading is equal
to or greater than 37.5 °C
(99.4 °F)
Memory: Storage of up to 30 measured
values
Dimensions: 165 mm x 40 mm x 20 mm
Weight: approx. 81 g, including battery
Operating
conditions:
Body mode: 16 - 40 °C
(60.8 - 104 °F),
Object mode:
5 - 40 °C (41 - 104°F),
Humidity: 15 - 95 % RH
Storage
conditions:
-20 to +50 °C (-4 to +122 °F),
Humidity: 15 - 95 % RF
Automatic
switch-off:
approx. 3 min after last
measurement
Quality guarantee:
The unit conforms to the Council Directive 93/42/
EEC concerningmedical devices, ASTM E1965
“Standard Speci cation for Infrared Thermo-
meters for Intermittent Determination of Patient
Temperature“ and draft standard IEC 80601-2-56
“Medical electrical equipment - Part 2-56: Par-
ticular requirements for basic safety and essential
performance of clinical thermometers for body
temperature measurement“. Geratherm
®
is certi-
ed in accordance with Council Directive 93/42/
EEC, Annex II and DIN EN ISO 13485 and is enti-
tled to af x the CE 0118 (Noti ed Body: LMET)
TECHNICAL DATA