Questions and Answers

  • T
    tonyawilcoxAug 20, 2025
    Why CM 100 XT, PhysioGate App ECG transmission (Bluetooth) does not function?
    • V
      vchanAug 20, 2025
      If the CardioMem CM 100 XT and smartphone are not coupled, check the INFO screen in the PhysioGate App for the serial number of the coupled device. If no or an incorrect serial number is displayed, start pairing. If the PhysioGate App does not run on the smartphone, unlock the screen and start the PhysioGate App. If the smartphone is not connected to the Internet, check the WWW status in the INFO screen in the PhysioGate App. If the status is not OK, change your location and try again.
  • A
    Alexis YoungAug 22, 2025
    What to do if CM 100 XT device cannot be turned on?
    • M
      michael21Aug 22, 2025
      If the CM 100 XT device cannot be turned on and the LED is not lit, it might be due to a low battery or no battery inserted. Insert a new battery to resolve this issue.
  • J
    John CarterAug 25, 2025
    What to do if CM 100 XT recording cannot be started?
    • H
      Holly AllisonAug 25, 2025
      If the CM 100 XT recording cannot be started and the LED is flashing orange, accompanied by 3 beeps when the pushbutton is pressed, it indicates an open lead. Check the contacts of the electrodes. If an electrode became detached, clean the skin and replace the electrode.
  • E
    Elizabeth SchwartzAug 26, 2025
    What to do if CM 100 XT recording cannot be started and the LED is lit orange?
    • J
      Jamie NguyenAug 27, 2025
      If the CM 100 XT recording cannot be started and the LED is lit orange, it means the memory is full. Connect the recorder to the PC and the CM 100 Configurator software and switch it on. Start CM 100 Configurator and check the recorder status. Download the ECG data and delete the memory of the recorder.
  • L
    Leslie BennettAug 28, 2025
    Why CM 100 XT LED is flashing red?
    • N
      Natalie RodgersAug 28, 2025
      If the LED on your CM 100 XT is flashing red, it indicates a low battery. Replace the battery of the CM 100 XT.
  • T
    Thomas StoneAug 29, 2025
    How to fix CM 100 Configurator error in the USB connection?
    • A
      Alyssa JacksonAug 30, 2025
      If you encounter an error in the USB connection with the CM 100 Configurator, and the message 'Error! USB device not connected!' appears, it could be that the USB download cable is not connected to the PC or the CM 100 XT is turned off. Connect the USB download cable to the PC. Turn on the recorder (the LED is lit orange after the starting sequence) and click on the button “Connect to recorder” to establish the connection for communication.
  • A
    Alex MorseAug 31, 2025
    What to do if CM 100 XT device does not switch to normal mode?
    • S
      stephen25Aug 31, 2025
      If the CM 100 XT device does not switch to normal mode and the LED is flashing alternately red, green and blue, it indicates a general device error. Remove and reinsert the battery. If the problem persists, contact your dealer or the service representative of the manufacturer.
  • L
    Luke WatsonSep 2, 2025
    How to fix CM 100 Configurator data download to a PC?
    • P
      pattersonkarenSep 2, 2025
      If the data download to a PC does not function in CM 100 Configurator and you receive the error message 'Error! No data download!', it means the ECG recordings could not be downloaded. Click on “Disconnect recorder” and then on “Connect to recorder” to reset the USB connection. If this does not function: Exit the CM 100 Configurator software. Remove the USB download cable (the device is turned off). Reconnect the USB download cable and restart the recorder. Start the CM 100 Configurator software. Click on “Connect to recorder” and repeat the download. Check that ECG reports are in the pre-set download folder.
  • T
    Thomas FowlerSep 4, 2025
    How to fix incomplete recording on CM 100 XT?
    • A
      Abigail WalshSep 4, 2025
      If you are experiencing incomplete recordings with your CM 100 XT, it may be due to a low battery. Replace the battery of the CM 100 XT.
  • R
    Randall StoutSep 6, 2025
    How to fix incorrect time stamp on getemed ECG report?
    • B
      Billy PerezSep 6, 2025
      If you notice an incorrect time stamp on the ECG report, it's likely due to an exhausted internal backup battery. Use the download cable to connect the CM 100 XT to a PC at least 12 hours to charge the backup battery.