Follow local standards if they are stricter that Getinge Disinfection
AB's recommendations, especially when treated water is used for
the final washing/disinfection phase. Typical specifications for
treated water is:
Parameter Unit Drinking
water
Softened DiW/
RO*
pH value N/A 6.5 - 8.5 6.5 - 8.5 5 - 7.5
Hardness mg/l < 120 < 10 N/A
Conductivity (at
25 °C)
µS/cm < 700 < 700 ≤ 15
Chlorine (Cl) mg/l < 60 < 60 ≤ 0.2
Iron mg/l < 0.3 < 0.3 ≤ 0.1
Copper mg/l < 0.1 < 0.1 ≤ 0.1
Silicon dioxide
(SiO
2
)
mg/l ≤ 2 ≤ 2 ≤ 1
Free chlorine mg/l < 0.5 < 0.5 ≤ 0.5
Phosphate mg/l < 10 < 10 ≤ 0.5
*DiW/RO = Treated water/Reversed osmosis
8 WATER QUALITY
46 (56) User Manual Getinge 46-series