G450 / G460 Baery pack performance and
maintenance procedures
TN2013_19_08_16
Tel: (800) 959-0329 or (734) 769-0573
Fax: (734) 769-1888
E-mail: info@gfg-inc.com
Website: www.gfg-inc.com
Interchangeable G450 and G460 rechargeable nickel-metal hydride
(NiMH) and alkaline baery packs are designed to provide up to 25
hours of connuous operaon. NiMH baeries provide excellent
cycle life and low temperature performance, and are warranted for
2-years. The typical G450 run-me aer two years with a properly
maintained NiMH baery pack is usually around 16 hours.
Some types of sensors require more power than others. The
highest power consuming sensors are photoionizaon detector
(PID) sensors, followed by standard catalyc LEL sensors and
infrared (IR) sensors. The power requirements for oxygen and
electrochemical toxic sensors (e.g. CO, H
2
S, SO
2
, NO
2
, etc.) are so
low that they do not materially aect the instrument run me.
The following chart shows the expected run-mes per charge (or
set of baeries) for various heavy-power sensor conguraons.
All conguraons include O
2
and CO/H
2
S sensors as well as the
listed “high power” sensors.
G50 and G460 rechargeable (NiMH) baery packs
provide outstanding operaon me, cycle life and low
temperature performance. Periodic deep discharge
ensures the best possible performance.
The temperature can also have an eect on the operaon me of
the instrument. While high temperatures up to 122°F (50°C) do
not have an effect on operation time, in very cold temperatures the
operaon me becomes shortened. Alkaline baeries should be
avoided when operang temperatures are below 32°F (0°C). NiMH
battery packs provide excellent low temperature performance
down to -4°F (-20°C), and can be used for short periods of me
down to -40°F (-40°C). The following chart shows the expected run
mes for a G450 equipped with a rechargeable NiMH baery pack
and standard LEL, O
2
, CO and H
2
S sensors in various temperature
condions: