Steps to verify and adjust the pump's flow rate to 2000 cc/min.
Instructions to perform after the pump has been charged.
Navigating pump menu to select and start calibration mode.
Connecting the IOM, calibration adapter, and tubing for calibration.
Attaching tubing to the calibration adapter and calibrator inlet.
Powering the calibrator and selecting the 'measure' option.
Configuring the calibrator to run continuously until stopped.
Connecting the second end of the tubing to the pump inlet.
Modifying the pump's flow rate to match the calibrator's average reading.
Stabilizing readings, recording the average flow, and calibration time.
Clearing the calibrator display and returning the pump to the main screen.
Connecting the calibrated pump and tubing to the person's belt.
Running tubing to the breathing zone and clipping the IOM filter correctly.
Turning on the pump and verifying it is in run mode with correct flow.
Pressing the check mark/power button and selecting 'Stop' to end the event.
Pressing the check mark/power button and selecting 'Pause' to temporarily stop.
Running the pump briefly and connecting to calibrator tubing for post-checks.
Allowing the piston to cycle and recording the average flow as the post-calibration reading.
Recognizing pump faults due to kinks or blockages, indicated by a red light.
Checking the fault time and details via the 'Review' tab on the main screen.