FAULTS EC 65
08/16 19
does not light or EC 65 cannot be switched on
This fault may be caused by:
1. No mains connection
Check whether the mains plug at the mains connection of the EC 65 and the shock-proof plug are
connected to the mains socket.
Make sure that the power supply at the mains socket is okay.
2. Mains fuse defective
The mains connection must be
opened, pull out the mains plug!
Switch off the EC 65 at the power switch (1) (press
0).
Pull out the main plug (2).
Pull out the fuse holder (3).
Check the main fuses (4).
Replace the mains fuse (see SPARES chapter for
part numbers).
Push in the fuse holder.
!
The fuse holder must snap in.
Figure 18 Check the mains fuse