765041976 Rev. 000 23
Door monitoring
When the door is opened, "OP" is shown in the primary display.
An acoustic signal is emitted, and the "A1" message is shown on
the secondary display, if the door at a minimum is open in "Dad",
and "BU" is connected.
The acoustic alarm is deleted by pressing
P
. The alarm indicator
only or first shuts off once the door has also been closed.
Cleaning the condenser filter
Reminder of cleaning the condenser air filter:
After 600 compressor running hours the filter must be cleaned and it is indicated by warning
lights:
FILTER .
If the cleaning is not completed within 650 hours, the warnings continue, and an acoustic
alarm sounds.
Resetting the FILTER alarm after cleaning the condenser filter:
After cleaning the air filter, the controller must be reset to remove the alarms.
It can only be reset by using a certain key combination.
Push three times followed by pushing three times.
FILTER alarm lights green
and will disappear after 1 minute.