English (GB)
8
7.5.1 "Last result"
Fig. 12 "Last result"
This submenu shows the data of the latest sample taken.
7.5.2 "View sensors"
Fig. 13 "View sensors"
This submenu shows the values of the different measurements.
7.5.3 "Current alarms"
Fig. 14 "Current alarms"
This submenu shows alarms related to the hardware. It shows an
E-number which can be looked up in section 8.2 Alarms and
warnings. The product will stop measuring when hardware alarms
are active.
Clearing alarms
1. Highlight the alarm you wish to clear and press [OK].
2. Highlight "Yes", press [OK] to clear the alarm or press [Right]
to choose "No" and [OK] to abort.
7.5.4 "Current warnings"
Fig. 15 "Current warnings"
This submenu shows warnings related to the hardware. It shows
an E-number which can be looked up in section 8.2 Alarms and
warnings.
7.5.5 "Current water quality alarms"
Fig. 16 "Current water quality alarms"
This submenu shows alarms related to the water quality. The
product will continue taking samples when water quality alarms
are active.
Some alarms may not be described in this manual, but they can
help Grundfos service specialists find the error in the product, if
needed.
Clearing alarms
1. Highlight the alarm you wish to clear and press [OK].
2. Highlight "Yes", press [OK] to clear the alarm or press [Right]
to choose "No" and [OK] to abort.
TM06 6405 2316TM06 6406 1416TM06 6407 2716
Alarms can be cleared manually; warnings cannot be
cleared manually. Warnings will disappear
automatically when the cause of the warning no
longer exists.
Home Service SettingsStatus
Sample ID: 29630
Analysis time: 2016-04-25 16:09:54
Bacteria concentration: 57430 /ml
Non-bacteria concentration: 98338 /ml
Flow cell: 91%
Measured volume: 6.13095 μl
Bad volume: 0.0192907 μl
Home Service SettingsStatus
Board temperature: 30 °C
Flow cell temperature: 20.59 °C
Flow cell pressure: 2.003 bar
Home Service SettingsStatus
2016-04-25 16:20:07 E41
Details about selected alarm:
Flow cell fouling is too high
TM06 6408 2716
Alarms can be cleared manually; warnings cannot be
cleared manually. Warnings will disappear
automatically when the cause of the warning no
longer exists.
TM06 6409 2316
Alarms can be cleared manually; warnings cannot be
cleared manually. Warnings will disappear
automatically when the cause of the warning no
longer exists.
Home Service SettingsStatus
2016-04-25 16:20:07 E41
Details about selected warning:
Flow cell fouling is too high
Home Service SettingsStatus
Bacteria concentration
Bad volume
Details about selected alarm: