Grason-Stadler GSI
®
61 CLINICAL AUDIOMETER
Status Messages
Using Default Data Indicates that the selected transducer and stimulus are not
calibrated to standard specified (ANSI/ISO) levels.
Erasing Data Indicates that the GSI 61 is clearing stored test information.
Saving Data Indicates that the test information is being saved into memory.
No Test Data Stored Indicates that there is no test data available to be erased or
printed.
Check Hi-Freq Cables Indicates that the High Frequency function is enabled but the
cables plugged into the L&R phone jacks are not the GSI-
specified type.
Invalid Selection Indicates that an incorrect selection, such as incompatible
transducers, has been made.
Not Available Indicates an uninstalled option has been selected.
No Test Data Stored Indicates that there is no data available for printing, although
the Print Audiogram button has been pressed.
Data Transfer Indicates that data has been processed either by the printer or
by a remote computer.
Check Printer A check is done to see if the printer is connected and ready to
receive data by checking the hardware flow control line. If
printer is not ready this message will be flashed.
Printing This is an informational message which is displayed while
outputting data to the printer. It appears on the lower right
section of the LCD.
No Printer Response If communications problems occur during the course of printing,
this error message will be flashed.
Communications Error Messages
The following communication errors may occur during data transmission through the RS 232
communications:
Comm error: parity Check GSI 61 and PC parity settings.
Comm error: framing Check GSI 61 and PC selections for number of stop
bits and number of bits per command.
Comm error: overrun See ‘framing’.
Comm error: multiple Check parity, baud rate, number of stop bits and
byte size on GSI 61.
Comm error: spurious Soft error. Should not be repeatable; if so, problem
could be anywhere in GSI 61 digital circuits or PCs
communication circuits.
Comm error: break See ‘spurious’.
If the word “Help” followed by an error code is displayed, note the Help Code number. See the
GSI 61 Service Manual (1761-0110) for detailed descriptions of these. Turn off the power to
the GSI61 and wait 30 seconds. Restart the instrument. Often the error will be cleared by this
action. If the error persists, contact an authorized Grason-Stadler service representative.
iii