EasyManua.ls Logo

Guardian 4 - Risks and Side Effects; Reagents; Removing the Sensor

Guardian 4
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 4 -
Risks and side effects
Risks related to sensor use include
Skin irritation or other reactions
Bruising
Discomfort
Redness
Bleeding
Pain
Rash
Infection
Raised bump
Appearance of a small “freckle-like” dot where the sensor was
inserted
Allergic reaction
Fainting secondary to anxiety or fear of needle insertion
Soreness or tenderness
Swelling at insertion site
Sensor fracture, breakage or damage
Minimal blood splatter associated with sensor needle removal
Residual redness associated with adhesive or tapes or both
Scarring
Reagents
The Guardian 4 sensor contains two biological reagents: glucose
oxidase, and human serum albumin (HSA). Glucose oxidase is
derived from Aspergillus niger and manufactured to meet industry
requirements for the extraction and purification of enzymes for use
in diagnostic, immunodiagnostic, and biotechnical applications.
The HSA used on the sensor consists of purified and dried albumin
fraction V derived from pasteurized human serum, which is cross-
linked via glutaraldehyde. Approximately 3 μg of glucose oxidase
and approximately 10 μg of HSA are used to manufacture each
sensor. HSA is approved for IV infusion in humans at quantities
much larger than in the sensor.
Removing the sensor
To change the Guardian 4 sensor, disconnect the transmitter from
the sensor as described in the Guardian 4 transmitter user guide.
Gently pull the sensor from the body to remove it. Discard the
sensor in a sharps container.