Pressurized portable eyewash supplies 0.65 GPM for 15 minutes, supplementing other emergency equipment.
Place unit where sprayheads are 33-45 inches high, no obstructions, on a flat, dry surface out of direct sunlight.
Using the unit for any purpose not intended by the manufacturer will void the warranty.
Read and follow all instructions to prevent serious eye damage and vision loss. Check pressure relief valve.
Inspect weekly for water level. Replace water immediately if particulation or discoloration is evident.
Depressurize tank using relief valve. Loosen bolts, nuts, clips, and clamp to remove eyewash assembly.
Fill tank with 10 gallons potable water. Add entire bottle of bacteriostatic additive. DO NOT OVERFILL.
Replace eyewash assembly, reclamp securely. Pressurize to 80 PSI using clean air source. DO NOT use CO2 or external sources.
Run unit momentarily to purge air. Drain and replace water/additive every six months. Keep unit away from heat.