EasyManua.ls Logo

Guildline 6675A - Page 140

Default Icon
162 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 7
7-10
7.2.3.2. Servo Loop Power Supply PCB Test Procedure
1. Using a DMM, measure the DC voltages at the Servo Loop PCB test points.
The voltages should be as shown in Table 7-4.
2. Using an oscilloscope, measure the noise voltage levels at the Servo Loop
Power Supply PCB test points. The noise levels should be as listed in Table
7-4.
Test points on PCB Assy 19107
(near J100)
Voltage
(V ±5%)
Noise Limits
(mV p-p)
TP313 with respect to TP301 -40 <100
TP310 with respect to TP301 +40 <100
TP303 with respect to TP301 -15 <100
TP300 with respect to TP301 +15 <100
TP311 with respect to TP301 +12 <100
TP314 with respect to TP301 +5 <500
J101-1 with respect to J101-2 +30 ±15% <800
Table 7-4 : Test Point Voltages Assembly 19107