EasyManua.ls Logo

Guildline 6675A - Page 159

Default Icon
162 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 7
7-29
29. Select the "High Volts" function by pressing the corresponding
menu key.
30. Wait 5 seconds (as per the displayed message) and observe the
returned "High Volts" reading.
31. Ensure the "High Volts" reading is stable over the next 10 readings
(± 1 count) (± 10 counts for 2 mV range).
32. Select
<Ok> by pressing the corresponding menu key.
33. Press the
<PREVIOUS> key until the
"Measure/Setup/Voltmeter/Sofcal" menu is displayed.
34. Select the "Voltmeter" function by pressing the corresponding menu
key.
35. Press the menu key under "20V Range" until the desired range
appears.
36. Ensure that the voltage displayed on the 6675A display is within the
voltage range indicated in Table 7-7.
37. Set the Precision DC Calibrator for -75% of full scale of the range
selected.
38. Ensure that the magnitude of the voltage displayed is within the
voltage range indicated in Table 7-7.
39. Set the Precision DC calibrator for +50% of full scale of the range
selected.
40. Ensure that the voltage displayed is within the voltage range
indicated in Table 7-7.
41. Set the Precision DC Calibrator for -50% of full scale of the range
selected.
42. Ensure that the magnitude of the voltage displayed is within the
voltage range indicated in Table 7-7.
43. Set the Precision DC calibrator for 0 V.
44. The voltage displayed should be very close to 0 V.