Section 3
3-18
are invoked for the four cycle precision measurements and are used in the calculation
of resistance ratio. Disabling the software filter function decreases the time required
to calculate and report the measured ratio. However, the data are now noisier, and
so two least significant digits are removed from the display when quick measure is
turned on.
Function key <Fn-4> returns to menu page 1 of 3.
3.7.2. <Setup> MENU Fn-2 (MAIN)
This menu function allows selection and edit of the measurement configuration.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒ ▒
▒ ▒
▒ ▒
▒Current Set Up ▒
▒ ▒
▒ ▒
▒══════════╤══════════╤═════════┬──────────▒
▒Load │Edit │Save │Resistors ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Probes
3.7.2.1. <Load> MENU Fn-1 (Setup_MAIN)
This menu function displays the list of configurations available. Up to 15 unique
configurations can be saved for review, 10 are shown on the first screen and 5 on the
second screen.
Each entry in the list is identified by the serial number entered for the Standard
Resistor in the test setup.
Where a temperature probe (one of up to five) is specified the serial number for that
probe is displayed at the list entry.