Section 3
3-34
3.7.3. <Voltmeter> MENU Fn-1 (MAIN)
The function keys behave the same as in the <Measure> menu and toggle through the
available options in the function key window.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
STATUS LINE
Voltmeter Reading
▒──────────┬──────────┬─────────┬──────────▒
20V Range Filter OffNormal MORE 1-2
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
2V Range Decimation Delta
200mV Rang N Average Delta SP
20mV Range
2mV Range
Function Key <Fn-1> ,<20V Range>, allows selection of the voltage range that the
voltmeter function will use during the test measurement.
Function Key <Fn-2> ,<Filter>, allows selection of the filter that will be used on the
displayed readings, where:
No Filter = display most recent reading
Decimation = display (100-N) percent of the previous reading.
Where N is the percentage value set in the
<Sofcal><Filter> menu for Decimation.
N Average = display the average of the last N readings. Where N
is the number of readings to be averaged and is that
value set in the <Sofcal><Filter> menu.
Function Key <Fn-3> ,<Normal>, allows selection of displaying the most recent reading
with no offset (<Normal>); or displaying the offset of the most recent reading from the time
that this softkey was last pressed (<Delta>); or displaying the offset of the most recent
reading from the value that was set in the <Sofcal><Datum> menu (<Delta SP>).