Section 3
3-41
Function Key <Fn-2>, <High Volts> takes a 5 second average of the reading
on the nanovolt meter in counts and displays it on the screen. Make sure that
the applied voltage is set between the P
2
terminals and that a voltage level is
entered for <Ref Volts>.
Function Key <Fn-3>, <Ref Volts> allows the actual value of the <High
Volts> to be entered.
Function Key <Fn-4>, <Ok> accepts all the information that has changed
and saves the changes to memory.
3.7.4.3.2.<Coeff Set> MENU Fn-4 (Voltmeter_Sofcal_MAIN)
This menu allows viewing and edit of the nanovoltmeter coefficients.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Zero Volts:
High Volts:
Ref Volts:
▒──────────┬──────────┬─────────┬──────────▒
Edit Ok
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Function key <Edit> allows editing of the coefficients displayed. Use the
arrow keys to move the highlight to the data line to be changed, modify the
data and press function key <Ok> to accept the value.