Section 3
3-42
3.7.4.4. <Diagnostic> MENU Fn-1 (Sofcal_MAIN)
Diagnostic menu selection page 1 of 4.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒Set Internal Values ▒
▒Reading: Range: ▒
▒Turns: <Fn-3> Mode: ▒
▒Sign: Speed: ▒
▒milliAmps: Cur Ramp: ▒
▒Volts: Volt Ramp: ▒
▒══════════╤══════════┬─────────┬──────────▒
▒Monitor │Dacs │Turns │MORE 1-4 ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Function Key <Fn-1>, <Monitor> selects the monitor menu level and displays the
value of the internal channels reported from the servo card assembly.
Function Key <Fn-2>, <Dacs> selects the values for a four channel 12-bit digital to
analog convertor on the servo card assembly.
Function key <Fn-3>, <Turns> allows the number of turns to be entered for the RX
side, from 0 8747 turns.
3.7.4.4.1.<Monitor> MENU Fn-1 (Diagnostic_Sofcal_MAIN)
This diagnostic menu allows monitoring of the signals on the servo board.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒Monitor readings on the Servo Board ▒
▒Servo: Prim. Cur: ▒
▒Speed Up: Calibrate: ▒
▒Offset: ▒
▒+40 Volts: ▒
▒-40 Volts: ▒
▒──────────┬──────────┬─────────┬──────────▒
▒ │ │ │ ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒