TRN-QRG-100125-US(AC)
This guide is not intended as a substitute for the Cell Saver
®
Elite
®
+ System Operation Manual or authorized training.
Load Disposable / Initiate Procedure (page 1 of 3)
Prepare the device:
1. Unfold biohazard waste bag, stored
in tray on underside of device.
2. Ensure biohazard waste bag is
connected to and hanging from
drain tube and open clamp.
3. Ensure pump platen lever and
device cover are closed and valve
module cover and centrifuge header
arm are closed and locked.
4. Power on device and allow
system self test to finish.
5. Select bowl size or scan
processing set when
prompted.
Load reservoir and A&A line:
1. Place reservoir in weigher and
close drain port slide clamp.
2. Connect vacuum line to device
vacuum port and yellow-
capped port on reservoir. If
using external suction, connect
external vacuum to yellow-
capped port on reservoir.
3. Receive and attach A&A line
to blue-capped port on
reservoir and insert into
tubing support.
4. Turn on suction. If using manual suction,
set to minimal acceptable level (<200
mmHg).
5. Close roller clamp on A&A line.
6. Hang and spike anticoagulant bag.
7. Prime reservoir until breakthrough
(150-200ml). Adjust roller clamp on
anticoagulant bag to 1 drop/sec.