TRN-QRG-100125-US(AC)
This guide is not intended as a substitute for the Cell Saver
®
Elite
®
+ System Operation Manual or authorized training.
Transition to Post-op
After surgery is complete:
Follow the steps below to transition the device to postoperative operation.
1. Turn off suction.
2. If necessary, connect filtered vacuum line to vacuum port on back of device and vacuum inlet port on reservoir.
3. Open Post-op suction set, receive, and attach Y-connector to wound drain tubing.
4. Disconnect A&A line and discard according to the facility's policies and procedures for handling of biohazardous materials.
5. Connect Post-op suction set to reservoir.
6. Power down device, disconnect from power, lower IV poles, and transport device to PACU.
NOTE: Do not select End Procedure.
7. Connect device to power, power on, and touch Resume Procedure.
8. Touch Suction and select Post-op suction from drop-down list (default: 75 mmHg).
NOTE: System will begin processing when minimum fluid level is collected.