What could falsely trigger the bowl optics sensor
into believing that the RBC layer is present when it
is not? (The RBC layer has not reached the sensor
height.)
§ Patient has a slightly lower than
average hematocrit
§ Supernatant that is perfectly clear
(close to pure water)
§ Supernatant that is dark (close to
RBC color) - YES
§ Dirty bulb or optics from a
previous spill - YES
No