Do you have a question about the Haier DW-86L628 and is the answer not in the manual?
Type | Upright Freezer |
---|---|
Temperature Range | -40°C to -86°C |
Power Supply | 220V/50Hz |
Defrosting | Manual |
Number of Shelves | 3 |
Color | White |
Capacity | 628 liters |
Refrigerant | Mixture Refrigerant |
Dimensions (H x W x D) | 1980 x 984 x 998 mm |
Provides information on backup cooling systems and associated warnings.
Details storage rack and box options for efficient internal space utilization.
Explains the 6-inch temperature recorder, its operation, and calibration.
Discusses the AC power supply and the function of the backup battery for the recorder.
Step-by-step guide for replacing the chart paper in the temperature recorder.
Details the procedure for calibrating the temperature recorder's accuracy.
Further details on the temperature recorder's operation and battery backup.
Explains the power supply requirements and the function of the backup battery.
Advises on inspecting and replacing the recorder's backup battery.
Illustrates the refrigeration system components and flow of refrigerant.
Shows the electrical connections for the freezer's control and power systems.
Specifies optimal ambient temperature, humidity, dust, and elevation conditions for installation.
Provides guidelines for selecting a suitable installation location, including ventilation and power requirements.
Identifies the battery location and outlines safe removal steps.
Lists potential causes and checks for when the freezer fails to start.
Covers common reasons for decreased cooling performance and checks.
Explains possible causes of unusual noise from the freezer and how to address them.
Instructions for removing packing materials and adjusting the leveling legs.
Emphasizes the importance of proper grounding and power connection for safety.
Details the procedure for initial startup, temperature setting, and performance testing.
Provides critical warnings and cautions regarding sample loading and unit care.
Explains how the freezer resumes operation after a power interruption.
Describes monthly cleaning procedures for the unit and condenser filter.
Provides instructions for safely defrosting the interior of the freezer.
Advises on replacing the battery for the power outage alarm due to its limited lifespan.
Explains how the buzzer alarm can be temporarily silenced and its automatic resumption.
Details the location and function of remote alarm terminals for external monitoring.
Step-by-step guide to unlock the control and adjust temperature set points.
Instructions for setting the high and low temperature alarm limits and their parameters.
Explains how to set and change the security code for unlocking the control panel.
Details the feature for delaying compressor startup to reduce power surge.
Describes how to perform functional tests for alarms and battery status.
Explains the meaning of various indicator lights on the control panel.
Lists various alarm conditions, their triggers, indicators, and buzzer sounds.