42
3
If the power cord needs to be lengthen, the cross section area of the extension line’s 
conductor should be at least 2 mm² and the length of the extension cord should be limited 
to 3 m.  This is to prevent electric fire or shock.
The power cord of the unit is equipped with a 16 A three-prone power plug. Do not remove 
the ground pin of the power plug under any circumstances.  Make sure the plug is securely 
plugged into the power outlet to prevent fire.
The power socket must be equipped with a ground wire to prevent electric shock.  If the 
socket fails to be grounded, the ground wire must be installed before the refrigerator is 
plugged in.
The refrigerator shall not conduct outdoor service. Electrical leakage or shock may be 
caused if wet by rainwater.
Do not place the refrigerator in humid places or places where the unit may get splashed 
on by water.   This is to avoid electric shock due to deterioration of insulation.
Dear users:
Thanks for your choice of Haier pharmaceutical refrigerator, please make sure you have carefully 
read and observed the contents with following signs in the manual, for better understanding of this 
manual and better use of this product, so as to prevent personal injuries and refrigerator damage.
In case the fire, do not pour water onto the refrigerator unit as a means to prevent electric 
shock or short circuit.
When there is a leak of flammable gas, shutoff the feed valve of the gas. Open windows 
for ventilation and exhaust.  Do not plug in your refrigerator or unplug your refrigerator as 
spark in these processes can cause an explosion or fire.
We recommend the unit be installed by a professional to avoid any electrical hazard.  
Place the refrigerator on solid and flat ground to avoid tipping the unit over to cause 
personal injury.
Only connect the refrigerator unit with a dedicated power outlet specified by the nameplate 
of the unit.  This is to avoid fire or electric shock.
If the supply voltage is lower than 198 V or higher than 242 V, an automatic voltage 
regulator of at least 4,000 W should be installed with the refrigerator.
Safety Precautions
Ignoring this warning may result in 
death or serious injury
Ignoring this warning may result in 
death or serious injury,and/or damage
to the refrigerator and property
Actions or operations 
which are prohibited
Actions or operations 
which must be followed
HYC-890
890L
R134a 250g
LED   6W
227kg
2℃~8℃
220-240V~/50Hz
800W/4A
Glass door with electric heat
Force-air cooling circulation
Cold-rolled steel with power coated/Cold-rolled steel with power coated
Pharmaceutical Refrigerator
Effective Volume
Door
Compressor
Shelves
Refrigerating Method
Refrigerant
Net Weight
Rated Power/Current
Rechargeable Battery
High & low temperature alarm, sensor error alarm, power failure alarm, 
low battery power alarm, door ajar alarm
DC 12V rechargeable battery
Name
Model
48 h (when the battery is fully charged)
Anti-shock Safety Classification
I
Power connection type
Y
Climate Type
4
Foaming Agent
CP/IP
Noise Level(Lp)
45dB(A)
1130×755×1980
1030×590×1425
Note: Climate Type 4 means the temperature is +30℃,the relative humidity is 55%.Technical information may vary from 
the appliance that you have just purchased due to technical improvements.
Exterior Dimensions
(W×D×H) 
(mm)
Interior Dimensions
(W×D×H) 
(mm)
Exterior / Interior 
Insulation
Load
Condenser/Evaporater
Interior light
Temperature Control
Temperature Range of 
sensor in glycerin bottle
Voltage
Alarm system
Battery duration for alarm system
CFC-Free  foamed-in-place urethane 
High Quality Hermetically sealed compressor
12 Polyester coated wire shelves
≤20kg per shelf
Microprocessor controlled
Specification
Fin tube type / Fin tube type