Control Panel
HYC-260/360/290/390/390F/890/940/940F
HYC-610
The recorder has been calibrated before leaving the factory.  Do not make adjustment except in 
special circumstances.
The recorder is optional for HYC-610 only, not available for others.
    Calibration of Recorder
The temperature recorder has been accurately adjusted before leaving the factory even if the power 
supply is interrupted.  If there is a need to be adjusted, perform the following procedures.
1. Let the unit run for at least 2 additional hours after it reaches stable condition,
2. Place a standard thermocouple in the temperature solution bottle filled with 100 ml 10% 
glycerol solution.  Place the sensor of the temperature recorder into the sensor bottle as well. 
3. Place the sensor bottle in the refrigerator for at least 4 hours.  Compare the thermocouple 
reading and the recorder reading.  If there is a discrepancy, press the adjustment button on the 
left (1#) or on the right (2#) of the panel to adjust the temperature value of the recorder so that 
recorder reading matches the reading of the thermocouple.  
Note:  The recorder pen will start moving 5 seconds after the button is left out.