Questions and Answers

  • A
    adamsglennAug 20, 2025
    What to do if I am unable to open the lid of my Haigh Quattro?
    • P
      Paige Jackson MDAug 20, 2025
      If you are unable to open the lid of your Haigh Medical Equipment, it may be due to a power failure. The lid cannot be opened until the power is restored and the interlock latch is released manually. Once open, check the operation of the latch and replace it if necessary.
  • H
    haydenjohnAug 22, 2025
    Why is there internal water leakage in my Haigh Medical Equipment?
    • S
      Steven SchneiderAug 22, 2025
      Internal water leakage in your Haigh Medical Equipment can occur if the Pulveriser or pump mechanical seal is leaking; in this case fit a new mechanical seal in the main assembly, or exchange the pump for a new part if that is where the leak is coming from. Also, investigate where the leakage originates and rectify the problem. Another potential issue is the Hopper seal; replace it, ensuring that the new seal fits correctly.
  • S
    Shannon OrtizAug 25, 2025
    How to fix no indicators illuminated on Haigh Medical Equipment?
    • M
      Matthew RamirezAug 25, 2025
      If no indicators are illuminated on your Haigh Medical Equipment, it could be due to a power failure. Check that the indicator membrane is connected properly. Also, check the fuses and electrical connections.
  • V
    Veronica WilkinsonAug 26, 2025
    What to do if my Haigh Quattro shows Drain Block Stage 2?
    • B
      Barbara GrayAug 27, 2025
      If your Haigh Medical Equipment shows a 'Drain Block Stage 2' error, there may be a blockage in the Pulveriser exit or drain. Investigate the cause and clear the blockage. Incorrect installation, pipe size, or position of waste pipework (less than 50mm minimum) could be the cause. Also, non-return/in-line valves may not be operating correctly; clean or replace them as required. Never use a reactive chemical drain block clearer within the machine, as it will damage the seals.
  • L
    Linda PenningtonAug 28, 2025
    How to clean the inside of my Haigh Medical Equipment if it's not clean after use?
    • B
      Barry WoodsAug 28, 2025
      If your Haigh Medical Equipment is not clean internally after use, check that the main water solenoid valve is operating and that the main isolating valve is open. Also, check that the water isolator service valve is in the ‘Open’ position and that the pump is working.
  • S
    Steven VargasAug 30, 2025
    How to stop a leaking lid seal on Haigh Medical Equipment?
    • R
      Randy BennettAug 30, 2025
      To address a leaking lid seal on your Haigh Medical Equipment, first check that the lid seal sits centrally on the lid gasket. Adjust and rectify this before making any adjustments to the latch mechanism. Also, remove any scum that has accumulated around the lid/hopper seal area.
  • L
    Laura JohnstonAug 31, 2025
    Why does the cistern on my Haigh Medical Equipment fail to replenish with water after 150 seconds?
    • Z
      Zachary LloydAug 31, 2025
      If the cistern on your Haigh Medical Equipment fails to replenish with water after 150 seconds, check that the mains inlet valve is open and that there is sufficient water pressure. Also, check and clean the inlet solenoid filter, and verify that the solenoid valve is operating; replace it if required.
  • M
    Mike BradleySep 2, 2025
    Why does the underside of the lid of my Haigh Quattro have a deposit after use?
    • G
      gregory81Sep 3, 2025
      If the underside of the lid of your Haigh Medical Equipment has a deposit after use, check for foreign matter in the gap of the vent in the centre of the lid. If this is blocked, remove the vent, clean it, and replace it. The jet gap must be the same around its circumference. If the problem is with the pump you will require an exchange unit for this part.
  • E
    Emily SullivanSep 4, 2025
    What to do if the water level in my Haigh Medical Equipment fails to drop after 20 seconds?
    • M
      Mary HopkinsSep 4, 2025
      If the water level in your Haigh Medical Equipment fails to drop after 20 seconds, check the pump and replace it if faulty. Also, check the cistern level switch and replace it if faulty.
  • M
    Mary AlvaradoSep 6, 2025
    What to do if I am unable to close the lid of my Haigh Quattro?
    • H
      Heather AndersonSep 6, 2025
      If you are unable to close the lid of your Haigh Medical Equipment, investigate and remove any foreign object present.

Summary

EC Declaration of Conformity

Safety Precautions & Procedures

Introduction and Safety Symbols

Manual usage overview, safety symbols explanation, and general instructions.

Operational Safety Guidelines

Covers electrical safety, staff training, dangers, and safe working practices.

Maintenance and Operation Safety

Specific safety for maintenance, modifications, and unacceptable operation.

Equipment Description, Process and Control

Machine Overview and Pulveriser

Details machine components, water supply, pulveriser function, and warnings.

Operation and Indicator Lights

Covers LCD panel, indicator lights, and operating procedures.

MaciFresh Kit Replacement

Step-by-step instructions for replacing the MaciFresh kit.

Specifications

Installation & Commissioning

Installation Requirements and Connections

Covers installation services, planning, water, waste, and overflow connections.

Electrical Setup and Commissioning

Details electrical, fusing, existing installation, and commissioning procedures.

Storage and Unpacking

Guides on storage, unpacking, and the installation procedure steps.

Maintenance & Servicing Procedures

Routine Maintenance Tasks

Daily, monthly, and quarterly checks for routine maintenance.

Cleaning, Lubrication, and Spares

Recommendations for cleaning, lubrication, and ordering spare parts.

Part Identification Diagrams

Assembly Diagrams and Parts Lists

Detailed diagrams and part lists for major assemblies.

Plumbing and Control System Diagrams

Diagrams for plumbing, control enclosure, wiring, and associated parts.

Optional Kits and Diagrams

Diagrams and details for optional kits like the foot pedal.

Fault Finding & Troubleshooting

Functional Problems and Solutions

Lists common functional issues and their possible causes/resolutions.

Troubleshooting Codes and Issues

Details troubleshooting codes (F1-F9) and specific problem resolutions.