WI-25023 19
PAIN MANAGEMENT SYSTEMS
Generator Service Manual
FAULT [H06]
Low Power Impedance
Measurement
Turn AC power OFF-ON to attempt
recovery.
1. Turn unit “OFF” then “ON” to
attempt recovery
2. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
FAULT [H07]
Corrupted Flash Disk
Turn AC power OFF-ON to attempt
recovery.
Table of Error, Fault & Warning Codes
Code Message
Recommended Action
(Complete List)
FAULT [H08]
Self-test failure: RF Current 2
Generation/Measurement
Turn AC power OFF-ON to attempt
recovery.
1. Turn AC power OFF-ON to attempt
recovery.
2. Allow 30-60 minutes with
generator in “ON” to acclimate the
unit to ambient temperature.
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
FAULT [H09]
Self-test failure: Cold Junction
Reference Error
Turn AC power OFF-ON to attempt
recovery.
Operational temperature may be
outside 10-40°C range.
FAULT [E10]
Internal Reference Calibration
Fault
Turn AC power OFF-ON to attempt
recovery.
Contact technical support if
problem persists.
1. Turn AC power OFF-ON to attempt
recovery.
2. Allow 30-60 minutes with
generator in “ON” to acclimate the
unit to ambient temperature.
3. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
FAULT [E11]
Duplicate variable storage
Turn AC power OFF-ON to attempt
recovery.
Contact technical support if
problem persists.
1. Turn AC power OFF-ON to attempt
recovery.
2. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
FAULT [E12]
Software Error Detected
Turn AC power OFF-ON to attempt
recovery.
Contact technical support if
problem persists.