EasyManua.ls Logo

Halyard PMG-115 Basic - Page 30

Halyard PMG-115 Basic
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WI-25023 30
PAIN MANAGEMENT SYSTEMS
Generator Service Manual
ERROR [E112]
Secondary Thermocouple
Connected But Disabled In
ADVANCED SETTINGS
(RFA/IDL)
1. Disconnect TDP B/Secondary
Thermocouple
2. If desired, enable TDP B/Secondary
Thermocouple in ADVANCED
SETTINGS
Table of Error, Fault & Warning Codes
Code Message
Recommended Action
(Complete List)
ERROR [E113]
Invalid RFA Temperature Reading
1. Check display settings
2. Check probe and cable
connections
3. Try new probe(s) and cable(s)
4. Contact HYH Technical Support
if problem persists (844-425-9273)
option 1 then option 3
ERROR [E114]
Invalid Secondary Thermocouple
Temperature Reading
Check probe and cable connections.
Probe or cable(s) may be defective.
1. Check Secondary Thermocouple
setting in Advance Settings
2. Check probe and cable
connections
3. Try new probe(s) and cable(s)
4. Contact Technical Support if
problem persists
ERROR [E115]
RFA and Secondary
Thermocouple Not Connected
Check probe and cable connections.
Probe(s) or cable(s) may be
defective.
ERROR [E116]
CRP B Connected But Disabled
In ADVANCED SETTINGS
Disconnect CRP B or, if desired,
enable 2 PROBES in ADVANCED
COOLED RF SETTINGS.
1. Check Secondary Thermocouple
setting in Advance Settings
2. Disconnect probe and attach to
other end of Y-cable (if applicable)
3. If desired, enable both probes in
ADVANCED COOLED RF SETTINGS,
4. Or, disconnect CRP B probe
5. Check probe and cable
connections
6. Try new probe(s) and cable(s)
7. Contact technical support if
problem persists
ERROR [E117]
CRP A Not Connected
Check probe and cable connections.
Probe or cable(s) may be defective.
ERROR [E118]
CRP B Not Connected
Check probe and cable connections.
If desired, enable 1 PROBE in
ADVANCED COOLED RF SETTINGS.
Probe or cable(s) may be defective.
ERROR [E119]
CRP A and CRP B Not Connected
Check probe and cable connections.
Probe(s) or cable(s) may be
defective.
ERROR [E120]
CRP A Connected To Wrong Side
Of Y-cable
Disconnect probe and attach to
other side of Y-cable.